Products
YG812H type protective clothing blood penetration tester
香港正版资料全年资料12323
香港正版资料全年资料12323

Scope of application

Used to test the resistance of medical protective clothing to penetration of synthetic blood under specified test conditions.

Guideline

GB 19082-2009 Technical requirements for medical disposable protective clothing 5.4.3 Resistance to penetration of synthetic blood

technical parameter

1. Test conditions: temperature is (20±2) , relative humidity is (30-80)%

2. Pressurization pressure: the pressure gauge shows the pressurization pressure, the pressurization pressure is adjustable, and the pressure adjustment rate is less than 3.5Kpa/s

3. Fixed clamp: a clamp that can produce 13.6Nm torque

4. Test pressure: 0Kpa, 13.8Kpa

5. Sample size: 75mm*75mm

6. Power supply: AC220V 50Hz 0.2KW

7. Dimensions: 550mm*450mm*750mm